Magnet ID :     Calibration
Directory : S:\magdata\LCLS-II\Phase Shifter\Calibration\126Gcm_+Bx_upstream\Final Results\Stretched Wire
   
Date : 06-29-2018
Time : 16:22:31 - 17:26:51
Operator : VV
   
Project : LCLS-II
Device Type : Phase Shifter
Run Number : 1
Data Type :
Run Comment : 126 Gauss*cm magnet rotated, +Bx, upstream pos
   
gap I1Y I2Y I1X I2X
-
-12.0 1.0 (5)
-3.5 0.2 (5)
121.1 0.8 (5)
40.6 0.6 (5)

integral gap -02.00;+00.00 -01.00;+00.00 +00.00;+00.00 +01.00;+00.00 +02.00;+00.00
I1Y -
-15.2 1.6 (3)
-12.1 0.5 (3)
-13.5 1.2 (3)
-12.1 0.5 (3)
-12.6 1.5 (3)
I2Y -
-3.8 0.5 (3)
-3.5 0.3 (3)
-3.8 0.5 (3)
-3.3 0.2 (3)
-3.5 0.3 (3)
I1X -
119.4 0.8 (3)
119.8 0.1 (2)
119.9 0.1 (2)
119.5 0.5 (3)
119.8 0.1 (2)
I2X -
40.4 0.0 (3)
40.4 0.4 (3)
40.5 0.2 (3)
40.2 0.1 (3)
40.4 0.4 (3)

integral gap +00.00;-02.00 +00.00;-01.00 +00.00;+00.00 +00.00;+01.00 +00.00;+02.00
I1Y -
-11.9 1.3 (3)
-11.8 1.3 (3)
-11.8 1.4 (3)
-12.3 0.5 (3)
-11.4 0.9 (3)
I2Y -
-3.2 0.3 (3)
-3.6 0.3 (3)
-3.5 0.3 (3)
-3.4 0.3 (3)
-3.6 0.3 (3)
I1X -
121.0 1.4 (3)
116.6 1.3 (3)
118.6 3.1 (3)
118.5 2.1 (3)
117.5 2.2 (3)
I2X -
40.6 0.3 (3)
40.3 0.2 (3)
40.5 0.3 (3)
40.4 0.2 (3)
40.4 0.4 (3)

   
Magnet ID :     Calibration
Directory : S:\magdata\LCLS-II\Phase Shifter\Calibration\126Gcm_+Bx_upstream\Final Results\Stretched Wire